Mobetron术中放疗加速器的物理测试和操作技术

陈毅, 沈文同, 赵胜光, 张毅斌, 段隆焱, 许赪

陈毅, 沈文同, 赵胜光, 张毅斌, 段隆焱, 许赪. Mobetron术中放疗加速器的物理测试和操作技术[J]. 实用临床医药杂志, 2016, (5): 88-91. DOI: 10.7619/jcmp.201605027
引用本文: 陈毅, 沈文同, 赵胜光, 张毅斌, 段隆焱, 许赪. Mobetron术中放疗加速器的物理测试和操作技术[J]. 实用临床医药杂志, 2016, (5): 88-91. DOI: 10.7619/jcmp.201605027
CHEN Yi, SHEN Wentong, ZHAO Shengguang, ZHANG Yibin, DUAN Longyan, XU Cheng. Physical testing and operation technique of Mobetron mobile linear accelerator for intra-operative radiation therapy[J]. Journal of Clinical Medicine in Practice, 2016, (5): 88-91. DOI: 10.7619/jcmp.201605027
Citation: CHEN Yi, SHEN Wentong, ZHAO Shengguang, ZHANG Yibin, DUAN Longyan, XU Cheng. Physical testing and operation technique of Mobetron mobile linear accelerator for intra-operative radiation therapy[J]. Journal of Clinical Medicine in Practice, 2016, (5): 88-91. DOI: 10.7619/jcmp.201605027

Mobetron术中放疗加速器的物理测试和操作技术

详细信息
  • 中图分类号: R815

Physical testing and operation technique of Mobetron mobile linear accelerator for intra-operative radiation therapy

  • 摘要: 目的 探讨Mobetron 1000移动式术中放疗加速器电子线的剂量学特性和操作技术.方法 对Mobetron加速器所配套的47个不同形状、尺寸的限光筒进行4、6、9、12 Mev共4种能量的物理数据测量分析,总结临床32例患者的术中放射治疗操作经验.结果 直径10 cm、端面0°的限光筒的4、6、9、12 Mev能量的最大深度剂量分别是0.6、1.2、1.6、2 em,治疗深度分别为0.9、1.8、2.5、3.4 cm.32例患者术中放射治疗剂量正确,操作顺利,无意外发生.结论 应用移动式术中放疗加速器具有正常组织保护好、治疗时间短、生物效应高等优势.
计量
  • 文章访问数:  248
  • HTML全文浏览量:  26
  • PDF下载量:  2
  • 被引次数: 0
出版历程

目录

    /

    返回文章
    返回
    x 关闭 永久关闭